Item
|
Parameter
|
Model
|
TOB-GB-4880-S
|
Glove Box
|
Dimensions:4880mm wide x 750mm deep x 900mm high,. Innersurface: stainless steel Outer surface: spray paint, white Material for Box:3mm thick. 304 stainless steel, acid resistant Windows -Slanted window, transparent tempered safety glass, thickness 8 mm, protective film. Glove Ports -Material is hard aluminum alloy, O-ring seal. Gloves - Butyl - 8" cuff diameter -0.4" Thick - 32" Length Dust Filter - 0.3 µm - 1 gas inlet and 1 gas outlet filter Shelves - Stainless -adjustable height. External Fluorescent Lights -LED lights, mounted in front of each glass window Interface: 16 spare interfaces, DN 40 KF. |
Main Antechamber
|
Large Antechamber - 304 Stainless Steel 360mm(14.2")Diam. - 600mm(23.6") Long (R or L) Surface: the inner surface is polished and the outer surface is painted (white) Sliding Tray on Stainless304 Door: double door, anodized aluminum material, thickness 10mm, vertical operation, with lifting mechanism Pressure gauge: analog display Control:Solenoid valve touch screen automatic operation |
Mini Antechamber
|
One - Mini Antechamber - 304 Stainless Steel 150mm(5.9") Diam. - 300mm(11.8") Long (R or L) Surface: the inner surface is brushed and the outer surface is painted (white) Door: double door with moving sliding plate, press-button opening method Pressure gauge: analog display Manual Operation via Hand Valve |
Gas Purification
|
Function:Removal of H2O and O2. Container material: 304 stainless steel. Purification material: copper catalyst: 9 kg; molecular sieve: 9kg Purification capacity: oxygen removal: 100L; water removal: 4Kg Water oxygen index: less than 1ppm |
Circulatory system
|
Working gas: nitrogen, argon, helium Cycle capacity: integrated fan flow 145m3 / h |
Regeneration
|
Operation: PLC automatically controls the regeneration process. Regeneration gas: mixed gas of working gas and hydrogen, (hydrogen 5-10%) |
Valve
|
Main valve: DN 40 KF, electric dynamic angle valve Control valve: electromagnetic integrated valve |
Control System
|
Function: including self-diagnosis, power-off self-starting feature, pressure control and adaptive function; automatic control, cycle control, password protection; unit control adopts Siemens PLC touch screen. Pressure control: The pressure working pressure of the control box and transition chamber can be set freely within +/- 10 mbar, and the automatic protection is beyond +/- 12 mbar. Pedal: Controls the pressure of the cabinet for easy operation and step-down. PLC (Programmable Logic Controller). |
Vacuum Pump
|
Specification: 12m3 / h, rotary vane pump, with oil mist filter, gas vibration control.
|
Display system
|
Using Siemens PLC touch screen smart 700, display operating status, cabinet pressure, system records, etc.
|
Vacuum system control situation
|
British EDWARDS vacuum pump, can be started manually or through PLC, flow rate 12m3 / h, can vacuum the transition chamber, and maintain the tank pressure balance, vacuum pump limit vacuum≤2x10-1pa |
Oxygen analyzer
|
Measuring range: 0 ~ 1000ppm The ZrO2 sensor avoids the problem that the fuel cell has a short life and cannot be exposed to the air |
Water analyzer
|
Measuring range: 0~500ppm platinum electrode structure (P2O5 structure, which can be cleaned to the factory level by simple cleaning). Wide range of applications, especially for lithium battery manufacturing and metal organic users, can restore the initial state through the cleaning regeneration program. Avoid the problem of scrapping once a pollution. |
Increase the organic solvent adsorber
|
Dimensions: diameter 238mm, height 407mm, material 3mm thick 304 stainless steel. Installed on the circulation line.
|